Key Specification |
Specification
Range
|
Resolution |
Accuracy |
A Route Sampling Flow |
28.3L/min |
|
|
B Route Sampling Flow |
28.3L/min |
|
|
Spray Flow |
(8~10)L/min |
|
|
Peristaltic Pump Flow |
(0.006~3.0)mL/min |
|
|
Front Pressure of A Route Sampling Flowmeter
|
(-20~0)kPa |
|
|
Front Pressure of B Route Sampling Flowmeter
|
(-20~0)kPa |
|
|
Front Pressure of Spray Flowmeter
|
(0~300)kPa |
|
|
Ambient Temperature |
(-40~99)℃ |
|
|
Negative Pressure of the Aerosol Chamber
|
(-90~-120)Pa |
|
|
Negative Pressure of the Cabinet
|
-50~-200pa |
Data Storage Ability
|
>100000 sets |
Vortex Mixer Test Tube Specification and Quantity
|
Φ16×150mm test tube,8pcs |
High Efficiency Particulate Air Filter Properties
|
Filter Efficiency≥99.99% for particles> 0.3μm |
Median diameter of aerosol generator mass
|
Average diameter: (3.0±0.3)μm; Geometric Standard Deviation≤1.5 |
Double Routes 6-lever Andersen Sampling apparatus
|
Level1>7μm; Level 2: (4.7~7)μm;
Level 3: (3.3~4.7)μm; Level 4: (2.1~3.3)μm;
Level 5: (1.1~2.1)μm; Level 6: (0.6~1.1)μm
|
|
600×85×3mm (Length×Diameter×Thickness) |
Total number of positive quality control sampler particles
|
(2200±500)cfu |
Negative Chamber Flow Speed
|
≥5m3/min |
Size of Negative Pressure Chamber Door(W×D)
|
1000×730mm |
Size of Main Machine(W×D×H)
|
1180×650×1300mm |
Size of Supporting Frame(W×D×H)
|
1180×650×600mm, Height within 10cm is adjustable
|
Noise of the Tester |
<65dB(A) |
Weight |
150kg |
Power Supply |
AC220V±10%, 50Hz |
Power Consumption |
<1500W |